Verified by manufacturer 2 months ago
Bespoke – Pro Model, Variable Air Volume
A fume cupboard designed to prevent explosive perchlorate salt formation.
Suitable for a range of research, pharmaceutical and military applications.
TION’s wash-down fume cupboard range is a ducted, high-capacity product designed to eliminate perchloric salt formation. The unit is equipped with a touch-screen control system and airflow alarms to maintain your staff’s safety no matter the circumstance.
Wash-down fume cabinets are fixed fume extraction systems fitted with angular walls and bespoke water spray nozzles. These nozzles wash chemical residues from behind the rear baffles and duct connections, eliminating the chance of explosive perchloric salt formation inside your laboratory.
Features and benefits:
Color | White |
Finish | Other |
Material | Polypropylene |
Shape | Rectangle |
Length | 1200 mm 1500 mm 1800 mm 2000 mm 2500 mm |
Width | 980 mm Bespoke |
Height | 2370 mm Bespoke |
Warranty description | One year |
Uniclass |
Standard | To BS EN 14175. |
Arrangement | Free standing unit. |
Special requirements | Manufacturer guidanceInsert requirements. |
Extract air | |
Face velocity | 0.5 ±0.1 m/s (100 ±20 FPM). |
Operation | Facility to vary extract volume flow rate. |
Materials | Polypropylene. |
Sash | |
Sash arrangement | Vertical.Horizontal.Vertical/ Horizontal. |
Operation | Powered, automatic. |
Maximum operational sash opening | 500 mm.Guidance for specification option: Front sash maximum opening 720 mm. |
Dimensions | |
Overall depth | Bespoke. |
Overall width | Bespoke. |
Lighting | |
Lighting type | LED lighting. |
Illuminance at work surface (minimum) | 800 lx. |
Services | |
Water | |
Sink unit | Not required.Required. |
Water supply | Not required.Cold water supply.Hot water supply.Reverse osmosis water supply.Demineralized water supply. |
Drainage | Manufacturer guidanceInsert requirements or omit. |
Electricity | |
Power requirements | 110 V, 50/60 Hz, single-phase.220 V, 50/60 Hz, single-phase.Manufacturer guidanceA three-phase power supply can be specially ordered. Consult manufacturer for details. |
Electrical socket outlets | Manufacturer guidanceSingle phase, 3A, 240 V, 50 Hz. Insert the required number of electrical socket outlets. |
Location | Sockets on front face of fume cupboard. |
Gas outlets | Not required.Nitrogen.Argon.Hydrogen.Vacuum.Compressed Air.Other.Guidance for specification option: Consult manufacturer and insert requirements. |
Airflow indicator with audible and visual alarms | Constant airflow monitoring and display with visual alarm on screen and integrated internal red light alarm system. |
Controls | |
Controller type | Touch screen control panel, 9-inch.Guidance for specification option: Full-colour touch screen with air flow velocity display, low airflow alarms display, fan control on/off, calibration settings, eco mode and VAV controls. |
Control modes | Eco mode. |
Accessories | Not required.Glove apertures.Multisash.Partitions.Shelves.Sumps.Vacuum tap. |
Worksurface | HPL. Ceramic. Epoxy. Polypropylene. Stainless steel. |
Safety devices | Not required. Co2/ powder fire suppression. External interlock emergency stop. |
Contains Red List materials | No |
Country of material origin | Germany |
Material | Polypropylene |
Expected Life | Ten year |
NBS BIM Object Standard
This BIM object has been authored to meet the requirements of the NBS BIM Object Standard.
Find out more about the NBS BIM Object Standard.
Platform | Compatible | Version |
---|---|---|
- | 1 | |
- | 1 |