NBS Source
I'm a manufacturer

Compare variants

Medipower™ Triple Cabinet IPS - Medical IT Power Supply
Medipower™ Triple Cabinet IPS - Medical IT Power Supply
Medipower™ Triple Cabinet IPS - Medical IT Power Supply

Verified 1 month ago

Brandon Medical Co Ltd

Verified 1 month ago

Brandon Medical Co Ltd

Verified 1 month ago

Brandon Medical Co Ltd
Summary
Isolated power supply for group 2 medical locations.
-
-
Application

Suitable for HD60364-7-710 Group 2 locations as well as HTM06-01 Category 5 areas.


Medical IPS systems should serve the following areas:

  • Operating theatre suites and recovery areas.
  • Emergency departments – resuscitation.
  • Critical care areas – CCU, ITU, SCBU, NICU, HDU etc.
  • Imaging rooms – MRI, CT, PET, X-Ray, Cath Lab etc.
-
-
Description

Medical IT Power Supply system (IPS) provides safe delivery of electrical power. The Medipower Medical IT Power Supply system (IPS) provides continuity of supply in the instance of the first insulation fault and also offers additional protection from electrical leakage currents, for group 2 locations or HTM06-01 for clinical risk category 5 areas.


Features and benefits:

  • High quality Medipower Medical IT Power Supply system (IPS) ensures safe energy distribution.
  • Distinct monitoring and distribution panel incorporates both wall-mounted and floor-standing options within the same panel design.
  • Automatic electrical supply, the standards state that Group 2 Medical Locations shall have an automatic electrical supply available within 0.5 seconds of power failure. In the UK, it is commonplace to support the IPS system with an upstream Uninterruptible Power Supply (UPS) in order to meet the 0.5-second requirement. In other parts of the world, this may be by switching to an alternative source of supply.
  • A CANbus (Controller Area Network bus) allows devices to communicate with each other without a host computer. IPS alarm signals communicate seamlessly via CANbus, offering many benefits.
  • CANbus is capable of identifying and reporting load insulation faults in 5–30 seconds.
  • Each CANbus network can be up to 2500 m in length for linking multiple networks or future expansion.
  • CANbus can offer over 50% reduction in field cabling costs.
  • CANbus continually monitors two-way communications between all CANbus connected devices, identifies any faults and provides parallel redundancy.
  • Modular construction and powerful messaging.
  • Ability to hot-swap and maintain individual components. Programmable content with up to 1000 alarm messages.
  • Smaller footprint space requirements.
  • IPS units can be supplied in as single (SIPS), double (DIPS) or triple (TIPS) configurations with a variety of different transformer sizes and in IPS, IPS+IFS or IPS+IFS+BATS variants.
  • Flexible number of outgoing ways - IPS system MCBs are available in functional blocks of six, up to a maximum of 96.
  • Multiple master alarm functionality.
  • Advanced event logging.
-
-
Product certifications
-
-
Digital objects
-
-
Color
Grey
-
-
Finish
Powder-coated
-
-
Material
Steel
-
-
Shape
Rectangle
-
-
Length
350 mm
-
-
Width
800 mm
-
-
Height
1700 mm
-
-

Specification

Standards
-
-
-
Generally
In accordance with BS 7671, HD 60364-7-710, and HTM 06-01.
-
-
Safety
To BS EN 60601-1 and BS EN 61558-2-15.
-
-
Monitoring
To BS EN IEC 61557-1, BS EN 61557-8 and BS EN 61557-9.
-
-
Service conditions
-
-
-
Treatment locations
Suitable for medical group 2.
-
-
Clinical risk
To HTM 06-01, category 5 areas.
-
-
Configuration
-
-
-
Isolating transformer
4 kV·A.
-
-
Distribution
-
-
-
Number of ways
Available in functional blocks of 6, up to a maximum of 96.
-
-
Surge protection
Not required., Required.
-
-
Automatic transfer
From primary to secondary within 0.5 s.
-
-
Provision for maintenance
Not required., Break before make bypass switch.
-
-
Communications
CANbus (Controller Area Network bus).
-
-
Integration with other systems
-
-
-
Secondary supply
Not required., Integrated uninterruptible power supply (UPS).
-
-
Enclosure
-
-
-
Arrangement
Modular.
-
-
Mounting type
Floor-mounted., Wall-mounted.
-
-
Material
Mild steel.
-
-
Finish
Powder-coated.
-
-
Colour
Light grey (RAL 7035).
-
-
Ingress protection (minimum)
IP41.
-
-
Ventilation arrangement
Three fans.
-
-
Ancillary products
Local alarm panels (LAP).
-
-

Additional properties

Configuration
Triple cabinet (TIPS).
-
-
Components
Incoming isolator.
Control and monitoring device - ILT.
-
-
Dimensions
1700 x 800 x 350 mm.
-
-
Air flow
42 m³/hr per IPS channel.
-
-
Protective earth bars
100 A.
-
-
Circuit breakers
24.
-
-
Rated frequency
50–60 Hz.
-
-
Input voltage
230 V a.c.
-
-
Current rating
17.4 A.
-
-
Power dissipation
Heat power dissipation no-load: <36 W.
Heat power dissipation 100%-load: <138 W.
-
-
Efficiency
96%.
-
-
Weight
52 kg.
-
-
Supply
Typical MCB Supply (A): C32 or D25.
Typical MCCB Supply (A): AF63/ AT30.
-
-
Accessories
Not required., MCB tripped monitoring., Network gateway., MODbus interface., Primary side outgoing RCBO’s (Type – 2C,25).
-
-

Sustainability data

Country of product manufacture
United Kingdom
-
-
Material
Steel
-
-