Verified by manufacturer 6 days ago
ATICS-2-63A-ISO, 63 A/ 240 V - B92057202
CE Marking
Certificate type
Automatic transfer switching devices with monitoring functions for unearthed safety power supplies.
Suitable for space-saving installations and retrofitting in Group 2 medical locations.
ATICS® transfer switching devices provide all functions for changeover between two independent power supplies and for monitoring unearthed power supplies. The power section and the electronic system integrated into one compact enclosure allow space-saving installation into the respective control cabinet, simplify wiring and reduce fault potential. To ensure maximum availability, ATICS® has been designed in strict accordance with the guidelines for functional safety.
Connectors at all connecting wires, in combination with an optional bypass switch, allow ATICS® to be tested without interruption of the power supply. During service works, it is possible to repair or replace the device without interrupting the power supply. ATICS® considerably enhances the safety level, particularly in intensive care units and in operating theatres.
Features and benefits:
For more information, please contact our team: internalsales@bender-uk.com.
Color | Yellow |
Material | Polycarbonate |
Shape | Other |
Length | 270 mm |
Width | 234 mm |
Height | 73 mm |
Uniclass |
Standards | To BS EN 60947-6-1, DIN VDE 0100-710, IEC 60364-7-710, IEC 61508 (SIL 2).Guidance for specification option: Refer to the data sheet for the full list of standards. |
Method of controlling transfer | ATSE. |
Operating sequence | I-I+II-II.Guidance for specification option: With Bypass switch set. |
Rated operational voltage (Ue) | 240 V. |
Rated insulation voltage (Ui) | 250 V. |
Rated operational current (Ie) | 63 A. |
Rated operational frequency | 50/60 Hz. |
Service conditions | Normal.Other. |
Terminals | Pluggable screw-type terminals. |
Controls | |
Type | Integral to switch. |
Display type | Liquid crystal display (LCD). |
Illuminated display to indicate the following conditions | Status indication of operating, warning and alarm messages. |
User-adjustable functions | Switching back interlocking function, manual/automatic mode, bypass operation, function test, changeover of the preferred line, alarm input for operating theatre lights and alarm input for other devices. |
Enclosure | |
Ingress protection (minimum) | To DIN EN 60529, IP40Guidance for specification option: LCD under foil. |
Material | Polycarbonate. |
Colour | Yellow and green. |
Accessories | None. B942709 - Measuring current transformer (load currentmonitoring IT system), Ie: a.c. 50 A. B98021000 - Measuring current transformer (short-circuit monitoring) for ATICS® < 100 A; Ie: a.c. 100 A. B98021002 - Measuring current transformer (load current monitoring IT system, Ie: a.c. 100 A. B92057252 - Bypass switch set (Ie: a.c. 63 A). Guidance for specification option: Comprising:
|